SEISMOLOGY AND GEOLOGY ›› 1982, Vol. 4 ›› Issue (4): 69-79.

• Brief Report • Previous Articles     Next Articles

THE PRESSURE CALIBRATION FOR ROCK TRIAXIAL TESTING APPARATUS WITH SOLID CONFINING MEDIA

Zhang Liu, Wang Shenzu   

  1. Institute of Geology, State Seismological Bureau
  • Online:1982-12-01 Published:2009-11-25

固体围压介质下岩石三轴实验装置的压力标定

张流, 王绳祖   

  1. 国家地震局地质研究所

Abstract: A study on pressure calibration of a triaxial testing apparatus with solid confining media is presented with temperature up to 625℃ and confining pressure up to 6.5kb. The solid confining media used were Pb,AgCl and Nacl. For room temperature, phase-transition materials, such as NH4I, phenol, AgI, NH4F and KNO2 were used to establish the relationship between the pressure applied on the specimen and the nominal pressurre measured outside The vessel. The correction of confining pressure △σc changing inversly with temperature was determined quantitatively by the "temperatueeffect" tests. It is shown that the pressure transmitting behavior of NaCl is obviously improved above 200℃ and a semi-emperical expression for △σc was derived for NaCl as a function of temperature T(℃), i. e. △σc = (0.491 + 0.000650T) [2.325-0.455 (T-25)0.217].